FOOD SCIENCE ›› 0, Vol. ›› Issue (): 219-222.

Previous Articles     Next Articles

Simultaneous Determination of Bromate, Chlorite, Chlorate and Common Inorganic Anions in Drinking Water by Ion Chromatography

  

  • Received:2011-06-10 Revised:2012-05-14 Online:2012-06-25 Published:2012-07-27

Abstract: An analytical method for the simultaneous determination of bromate, chlorite, chlorate and common inorganic anions in drinking water was developed using large-volume direct injection and chemically suppressed ion chromatography (IC). An IonPac AS9-HC separation column and an IonPac AG9-HC protection column were used. The mobile phase was 12 mmol/L Na2CO3 at a flow rate of 1.0 mL/min. The injection volume was 100μL. Ten anions could be detected by the method within 18 min with average spike recovery of 93.0%~103.1% (n = 5). The correlation coefficients of calibration curve were 0.9993~0.9999. This method was sensitive, accurate and rapid. It has been applied to determine Kunming tap water and a variety of commercial drinking water with satisfying results.

Key words: ion chromatography, drinking water, bromate, inorganic anions