FOOD SCIENCE ›› 2008, Vol. 29 ›› Issue (7): 332-335.

Previous Articles     Next Articles

Flow Injection Analysis-Spectrophotometric Determination of Soluble Protein Content

 XIAO  Kai, FENG  Yuan-Chun, ZHANG  Xin-Shen   

  1. College of Light Industry, Textile and Food, Sichuan University, Chengdu 610065, China
  • Online:2008-07-15 Published:2011-07-28

Abstract:  A method was developed for determining soluble protein by flow injection-spectrophotometric analysis with bromocresol green as color developing reagent and 630 nm as detecting wavelengh. Under the optimum conditions, working curve with good linear was obtained. The linear range of BSA is 1.0 ~280 mg/L, the detection limit is 0.52 mg/L, and relative standard deviation (RSD) is 2.26%. The range of recovery rates for different spiked samples is between 96.1% and 104.2%. The characteristics of this method are simple, rapid, and high sensitive and also can be used on-line.

Key words: flow injection analysis, spectrophotometry, bromocresol green, protein