Abstract: An atmoic fluorescence spectrometry for the assay of micro arsenic content in tea was developed with high sensitivity and fine accuracy.The curve range was 0~180ng/ml with correlation coefficient 0.9993.The limit of detection was as low as 4μg/kg and the recovery rate was 90.0%~97.3%.