Methodology Validation for Determination of L-Zinc Lactate by High Performance Liquid Chromatography with A Chiral Column

HE Yueming1, WANG Shan2, JIE Qinfeng2, QIU Weihua2, HUANG Xiaowan1, DENG Zeyuan1,*   

  1. 1. State Key Laboratory of Food Science and Technology, Nanchang University, Nanchang 330047, China;
    2. Center for Disease Control and Prevention of Jiangxi Province, Nanchang 330029, China
  • Online:2015-07-25 Published:2015-07-15
  • Contact: DENG Zeyuan

Abstract:

This paper describes the development of a validated method for determining L-zinc lactate by high performance
liquid chromatography (HPLC). An ideal chromatographic separation between D-zinc lactate and L-zinc lactate was achieved
on an OA-5000 chiral column using 1 mmol/L CuSO4·5H2O as mobile phase at a flow rate of 1 mL/min. The established
calibration curve showed a good linearity over the concentration range of 0.147 to 3.675 mg/mL, with a correlation
coefficient of 0.999 9 (n = 6). The relative standard deviation (RSD) for precision and stability were respectively 0.044%
was and 0.061%, and the average recovery was (99.8 ± 1.90)% with RSD of 1.90%. It is shown that this method is simple,
accurate and reliable, and can be used to monitor the content of L-zinc lactate.

Key words: high performance liquid chromatography (HPLC), L-zinc lactate, chirality