FOOD SCIENCE ›› 2011, Vol. 32 ›› Issue (18): 278-281.doi: 10.7506/spkx1002-6630-201118062

• Analysis & Detection • Previous Articles     Next Articles

Determination of Manganese in Edible Oil by GF-AAS with Direct Injection of Microemulsion

CHEN Shang-long,LIU Quan-de*,LI Chao,LI Xiang-hua   

  1. (College of Food Engineering, Xuzhou Institute of Technology, Xuzhou 221008, China)
  • Online:2011-09-25 Published:2011-09-24

Abstract: Graphite furnace atomic absorption spectrophotometry (GF-AAS) coupled with direct injection was used to establish a determination method for manganese in edible oil. The formation conditions of microemulsion and the effects of C12H25NaO4S concentration, ashing temperature and atomization temperature on absorbance were explored. Based on one-factor-at-a-time experiments, the optimal determination conditions were determined by orthogonal array design to be C12H25NaO4S concentration of 4.0 mg/mL, ashing temperature of 900 ℃ and atomization temperature of 1900 ℃. Under the optimal conditions, the content of manganese in edible oil was determined to be 37.49 ng/mL with a relative standard deviation of 1.86% and the detection limit of the method was 0.88 ng/mL. The average recovery rate for manganese in edible oil across three spike concentrations was 101.8% with a relative standard deviation of 3.27%. Therefore, this method is rapid, accurate and applicable without pollution and complex sample pretreatments.

Key words: microemulsion, direct injection, graphite furnace atomic absorption spectrophotometry, edible oil, manganese

CLC Number: