Spectrophotometric Determination of Chlorine Dioxide in Tap Water Using Methyl Red

CHEN Mei-mei,GAO Mei,DONG Han,LI Fen-yue,LIU Xiao-geng*   

  1. (Nanjing University of Finance and Economics, Nanjing 210003, China)
  • Online:2011-10-25 Published:2011-10-12

摘要: 探讨以二氧化氯(ClO2)氧化甲基红(MR)褪色光度法测定自来水中ClO2的新方法。在c(MR):c(ClO2)=6.4:1、pH4.0~4.5、温度25℃、反应时间40min条件下,以λmax=518nm测定δA518。结果其线性方程为δA518=-1.7845c(ClO2)+1.5228(r=0.992),表观摩尔吸光系数ε=1.2×105L/(mol ·cm),适用测定范围为0.0004~0.80mg/L,DL为0.0036mg/L;ClO2氧化MR褪色反应具有一级反应的特点,在25℃的动力学方程为ln[c∞/(c∞-c)]=-0.000689t+2.263669、Ea为87.3kJ/mol、t1/2为16.8min。本法经自来水和消毒剂中ClO2含量的测定结果表明,相对标准偏差为2.43%~3.83%,加标回收率为98.2%~103.2%。经F检验和t检验,本法与对氨基-N,N-二乙基苯胺(DPD)光度法不存在显著差异(P=0.05),测定结果符合分析的要求。

关键词: 自来水, 杀菌消毒剂, 二氧化氯, 甲基红, 褪色光度法

Abstract: A new spectrophotometric method for the determination of chlorine dioxide (ClO2) in tap water was developed based on methyl red (MR) oxidation and discoloration. Wavelength scanning showed that MR had maximal absorption at 518 nm. Under the experiment conditions of c(MR)/c(ClO2) = 6.4, pH 4.0-4.5, 25 ℃ and reaction time of 40 min, a linear equation was obtained as δA518 = -1.7845c(ClO2) +1.5228 (r = 0.992) in the range of 0.0004-0.80 mg/L, the apparent molar absorptivityε = 1.2 × 105 L/(mol ·cm), and the detection of limit 0.0036 mg/L. MR discoloration by ClO2 oxidation was a first-order reaction. The dynamic equation at 25 ℃ was ln[c∞/(c∞-c)] =-0.000689t + 2.263669 with the activation energy Ea = 87.3 kJ/mol and the half-time t1/2 = 16.8 min. The average recoveries for ClO2 in tap water and ClO2 disinfectant were 98.2%-103.2% (n = 5) with a relative standard deviation of 2.43%-3.83%. This method and the recommended standard N,N-diethyl-p-phenylendiamin (DPD) method showed no significant difference (P = 0.05) according to F-test and t-test. Furthermore, satisfying determination results were obtained from this method.

Key words: tap water, disinfectant, chlorine dioxide, methyl red, discoloration, spectrophotometry