FOOD SCIENCE ›› 2010, Vol. 31 ›› Issue (8): 226-229.doi: 10.7506/spkx1002-6300-201008051

• Analysis & Detection • Previous Articles     Next Articles

Simultaneous HPLC Determination with ELSD of Five Sugars in Foods

LUO Jin1,XIA Min2,YE Neng-sheng1,GU Xue-xin1,*,JIA Li2,CAO Ying-Hua2   

  1. 1. Department of Chemistry, Capital Normal University, Beijing 100048, China;
    2. Beijing Center for Physical and Chemical Analysis, Beijing 100089, China
  • Received:2009-06-30 Revised:2009-10-26 Online:2010-04-15 Published:2010-12-29
  • Contact: GU Xue-xin1 E-mail:guxuexin@263.net

Abstract:

A method for the simultaneous determination of 5 sugars such as fructose, glucose, sucrose, maltose and lactose in foods is described, in which high performance liquid chromatography (HPLC) and evaporative light scattering detection (ELSD) are used. A waters-NH2(4.6 mm × 250 mm, 5 μm) column kept at 30 ℃ and eluted with 75% acetonitrile aqueous solution at a flow rate of 1.0 mL/min was used for the chromatographic separation. The detector drift tube was kept at 45 ℃ and the nitrogen pressure was set at 25 psi. A good separation among the analytes and other components was achieved within 20 min (R > 1.5). The response of the detector to analyte concentration over the range from 50 to 2000μg/mL displayed good linearity. Precision RSDs of 6 replicates for the determination of the sugars were less than 3.8% and mean recoveries for the sugars in 4 real samples spiked at the levels of 100, 200 μg/mL and 500μg/mL were more than 80.3%.

Key words: HPLC, ELSD, food, sugar