FOOD SCIENCE ›› 2008, Vol. 29 ›› Issue (11): 498-500.

Previous Articles     Next Articles

Determination of Aluminum in Food with Tantalum-coated Graphite Tube-Graphite Furnace Atomic Absorption Spectrometry

 ZHANG  Yan, 吕Pin , LI  Hui, WANG  Duo, LIU  Jing-Ze   

  1. 1.College of Life Science,Hebei Normal University,Shijiazhuang 050016,China;2.Hebei Key Laboratory of Food Safety,Shijiazhuang 050051,China;3.College of Basic Medical,Hebei Medical University,Shijiazhuang 050017,China
  • Online:2008-11-15 Published:2011-12-08

Abstract: Aluminum in the wet-digested food was determined by tantalum-coated graphite tube-graphite furnace atomic absorption spectrometry with matrix modifier.Results showed that the linear range is between 0 and 50 μg/L with r =0.9996,RSD of 2.98% and recovery rate of 98.6% to 100.5%.The detection limit of the method is 0.27 pg.The sample analysis results measure up.This method is simple,rapid and accurate with satisfactory results.

Key words: tantalum-coated graphite tube, graphite furnace atomic absorption spectrometry, food, aluminum :