FOOD SCIENCE ›› 2008, Vol. 29 ›› Issue (3): 415-417.

Previous Articles     Next Articles

Assay Taurine in Milk Power for Infant and Young Children with Pre-column Derivatization HPLC

 HUANG  Zhe-Wen, TANG  Qin-Qiang   

  1. Shenzhen Neptunus Biology and Technologic Co. Ltd., Shenzhen 518057, China
  • Online:2008-03-15 Published:2011-08-24

Abstract: Objective: Strength the reliability of detemining the assay for taurine in milk power for infant and young children with pre-column derivatization HPLC. Methods: HPLC method was used, chromatography condition included:The instrument was SHIMADZU-10A with a Kromasil C18 column, and the mobile phase was methanol-0.05mol/L NaAc (35:65) at flow rate of 1.0 ml/min. The UV detection wavelenge was 360 nm. Results: Taurine had a good linearity(γ=0.9996)in the range of 0.00~20.34μg/ ml. The relative recovery was 99.6%, RSD was 2.0%(n=5). Conclusion: An accurate method with good reproducibility for control of taurine in milk power for infant and young children were established.

Key words: taurine, assay, HPLC, milk power for infant and young children