FOOD SCIENCE ›› 2008, Vol. 29 ›› Issue (12): 531-532.

Previous Articles     Next Articles

Determination of Gemanium and Selenium in Ganoderma iucidum Spore Powder by Using ICP-MS with Microwave Digestion

 LIU  Jun, CHEN  Jian-Ping   

  1. Guangdong Key Laboratory of Chemical Emergency Test,China National Analysis Center,Guangzhou 510070,China
  • Online:2008-12-15 Published:2011-12-08

Abstract: The method of determination for germanium and the selenium in the Ganoderma lucidum spore powder by means of the inductive coupling plasma mass spectrum(ICP-MS) with microwave digestion was established,taking 115In as internal standard. The limited of detection of germanium is 0.0003 mg/L(3δ),and that of selenium is 0.0005 mg/L(3δ). The standard curve range of germanium is 0.0005 to 0.005 mg/L(r=0.999 9),and that of selenium is 0.001 to 0.25 mg/L(r=0.9999). The relative standard deviation of germanium is 3.8%(n=10),and that of selenium is 1.7%(n=10). The recovery of germanium is 94% to 105%,and that of selenium is 93% to 104%.

Key words: ICP-MS, spore powder, germanium, selenium